Detect hazardous reactions and optimise your process design.
CONTACT USOur UK laboratory generates thermal screening and Differential Scanning Calorimetry (DSC) test data for your hazardous materials and chemical reactions.
The Sigma-HSE Advantage
Request a Free DSC Quote
Frequently asked questions
What is your standard methodology?
ASTM E 537-12
What is the Differential Scanning Calorimetry test?
The Differential Scanning Calorimeter (DSC) is a widely used thermal analysis instrument and is conducted as a screening method to assess the extent and presence of enthalpic changes and to approximate both the onset temperatures and decomposition enthalpies (heats) (heats) of solids, liquids or slurries. Very small sample sizes of 5-10 mg are used, complimenting lab-stage process developments for hazard assessment.
Qualitative information regarding material vaporisation and melting onsets also help to assess reaction hazards over a large temperature range.
The DSC test can detect potentially hazardous reactions from volatile chemicals while estimating the probability of a hazard and the time at which a hazardous reaction may occur.
How does the Differential Scanning Calorimetry test work?
This test is conducted by inserting the desired material and a thermally inert reference material into isolated crucibles within the instrument. Both the testing sample and the reference material are then simultaneously heated at a controlled rate of 2 °C/min to 20 °C/min under an equilibrated atmosphere. As the furnace temperature ramps up at the desired rate, the temperature difference between the sample and reference are recorded as heat flow (Δq) and is associated with changes in enthalpy of the sample.
What data can I get from the Differential Scanning Calorimetry test?
Thermal transitions in state (solid-liquid, liquid-vapour, glass transitions) typically come with an endothermic heat flow as the sample takes on energy to achieve a higher energy state, whereas thermal decompositions typically involve chemical bond breaking and heat release as exothermic energy. These exothermic events pose a key hazard in processes.
Characteristics such as Specific Heat Capacity can also be elucidated from DSC analysis.
Analysis of the data is performed to indicate possible hazardous events in the context of your processes.
What are benefits of the Differential Scanning Calorimetry test?
The DSC test can detect potentially hazardous reactions from volatile chemicals while estimating the severity of a hazard and the conditions in which a hazardous reaction may occur. The DSC can also be used as an early-stage test for detecting the potentially thermally hazardous nature of uncharacterised materials.
The DSC test is often run as a screening test complimentary to Carius Tube test, as it provides the valuable Heat of reaction data calculated from temperature and heat flow not available from the Carius Tube.